Severity
Moderate
FDA Devices recall · Reported April 11, 2018
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Addit…
Baxter Healthcare Corporation recalled Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate sy… - a moderate-severity action.
Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate sy… was recalled by Baxter Healthcare Corporation in April 11, 2018. Reason: Firm has received reports of device operators failing to adhere to instructions for use pertaining to the saf…. Check the official notice for the remedy. Verify recall #Z-1280-2018 with the FDA Devices before acting.
The recall
Baxter Healthcare Corporation issued this moderate-severity FDA Devices recall-Firm has received reports of device operators failing to adhere to instructions for use pertaining to the saf….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1280-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1280-2018) was formally reported on April 11, 2018, with the manufacturer initiating the action on February 15, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Baxter Healthcare Corporation is listed as the recalling firm, operating out of Deerfield, IL. Federal records list the affected scope as 3,440 units.
The documented reason for this recall is: Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and … Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Algeria, Armenia, Australia, Bahrain, Belarus, Belgium, Canada, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, I…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3,440 units
Related Recalls
6
3 from same agency
Prismaflex Control Unit. Dialyzer, high permeability with or without sealed dialysate system. Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. *Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated.
Firm has received reports of device operators failing to adhere to instructions for use pertaining to the safe unloading of disposable sets from the Prismaflex Control Unit. Additionally, for software versions 5.10 and 6.10, the programmed syringe size for the syringe pump may revert to safe default values unintentionally.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1280-2018 |
| Date reported | April 11, 2018 |
| Date initiated | February 15, 2017 |
| Recalling firm | Baxter Healthcare Corporation |
| Firm location | Deerfield, IL |
| Affected scope | 3,440 units |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Algeria, Armenia, Australia, Bahrain, Belarus, Belgium, Canada, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, Ireland, Italy, Alban… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 11, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.