Severity
Moderate
FDA Devices recall · Reported April 11, 2018
Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic …
Invivo Corporation recalled DynaCad software consist of an MR Analysis Server software and the viewer workstation sof… - a moderate-severity action.
DynaCad software consist of an MR Analysis Server software and the viewer workstation sof… was recalled by Invivo Corporation in April 11, 2018. Reason: Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered co…. Check the official notice for the remedy. Verify recall #Z-1289-2018 with the FDA Devices before acting.
The recall
Invivo Corporation issued this moderate-severity FDA Devices recall-Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered co….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1289-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1289-2018) was formally reported on April 11, 2018, with the manufacturer initiating the action on January 24, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Invivo Corporation is listed as the recalling firm, operating out of Gainesville, FL. Federal records list the affected scope as 470 distribution events, 232 actual units (updated 6/26/18).
The documented reason for this recall is: Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly w… Distribution data in the federal record shows the product reached: Device is software. Customer notification letters recommended that users discontinue use of the Ktrans, Kep, and iAUGC colormaps when assessing studies from a remotely connected DynaCAD client computer. It is further a…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
470 distribution events, 232 actual units (updated 6/26/18)
Related Recalls
6
3 from same agency
DynaCad software consist of an MR Analysis Server software and the viewer workstation software. The MR Analysis software consists of DynaCad Breast, DynaCad Prostate, and DynaCad Advanced PK for other MR analyses modules. Product Usage: Intended to be used as a post processing software package designed to provide a reliable means for analyzing MR datasets.
Following update from 3.3 to 3.5 of the DynaCAD software it was noted that the Ktrans map was not rendered correctly on the DynaCad Client. The defect causes the pharmacokinetic (PK) color maps to display incorrectly when viewed from remote DynaCAD client computers and could result in visually underestimating calculated Ktrans, Kep, and iAUGC values. The defect also impacts DynaCAD s on-the-fly calculation of Apparent Diffusion Coefficient (ADC) maps. If the ADC values are computed on-the-fly by DynaCAD, the ADC values and colors will also display incorrectly if viewed on a remote DynaCAD client.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1289-2018 |
| Date reported | April 11, 2018 |
| Date initiated | January 24, 2018 |
| Recalling firm | Invivo Corporation |
| Firm location | Gainesville, FL |
| Affected scope | 470 distribution events, 232 actual units (updated 6/26/18) |
| Distribution | Device is software. Customer notification letters recommended that users discontinue use of the Ktrans, Kep, and iAUGC colormaps when assessing studies from a remotely connected DynaCAD client computer. It is further advised that users re… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 11, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.