FIBERNET Embolic Protection System: Recommended Vessel Diameter 5.0 -6.0 mm, Maximum Guidewire Diameter: 0.014 in (0.36mm) Rx only, Sterile; Manufactured in: Invatec S.P.A., Italy; Manufacturer: Medtronic Inc. Minneapolis, MN. The FiberNet Embolic Protection System is indicated for use as a guidewire and emboli protection system to capture and remove embolic material (thrombus/debris) produced while performing percutaneous transluminal interventional procedures in carotid arteries in

Reported: April 2, 2014 Initiated: March 3, 2014 #Z-1304-2014

Product Description

FIBERNET Embolic Protection System: Recommended Vessel Diameter 5.0 -6.0 mm, Maximum Guidewire Diameter: 0.014 in (0.36mm) Rx only, Sterile; Manufactured in: Invatec S.P.A., Italy; Manufacturer: Medtronic Inc. Minneapolis, MN. The FiberNet Embolic Protection System is indicated for use as a guidewire and emboli protection system to capture and remove embolic material (thrombus/debris) produced while performing percutaneous transluminal interventional procedures in carotid arteries in high surgical risk patients with reference vessel diameters of 3.5 mm to 7.0 mm.

Reason for Recall

Medtronic is recalling a specific subset of lot numbers of the FIBERNET Embolic Protection System (EPS) where the yellow polyimide tube on the FIBERNET EPS was not removed from the FIBERNET EPS during the manufacturing process and subsequently may remain on the FIBERNET EPS.

Details

Recalling Firm
Medtronic Vascular, Inc.
Units Affected
50 Total all sizes: 18 in US, 32 Internationally.
Distribution
Worldwide Distribution - US including the states of IL, PA, MA, VA. RI and LA., and the countries of Germany, Italy, Poland, Romania, Saudi Arabia, Kosovo, Slovakia and Spain.
Location
Santa Rosa, CA

Frequently Asked Questions

What product was recalled?
FIBERNET Embolic Protection System: Recommended Vessel Diameter 5.0 -6.0 mm, Maximum Guidewire Diameter: 0.014 in (0.36mm) Rx only, Sterile; Manufactured in: Invatec S.P.A., Italy; Manufacturer: Medtronic Inc. Minneapolis, MN. The FiberNet Embolic Protection System is indicated for use as a guidewire and emboli protection system to capture and remove embolic material (thrombus/debris) produced while performing percutaneous transluminal interventional procedures in carotid arteries in high surgical risk patients with reference vessel diameters of 3.5 mm to 7.0 mm.. Recalled by Medtronic Vascular, Inc.. Units affected: 50 Total all sizes: 18 in US, 32 Internationally..
Why was this product recalled?
Medtronic is recalling a specific subset of lot numbers of the FIBERNET Embolic Protection System (EPS) where the yellow polyimide tube on the FIBERNET EPS was not removed from the FIBERNET EPS during the manufacturing process and subsequently may remain on the FIBERNET EPS.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 2, 2014. Severity: Moderate. Recall number: Z-1304-2014.