Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.

Reported: April 7, 2021 Initiated: March 3, 2021 #Z-1313-2021

Product Description

Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.

Reason for Recall

An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.

Details

Units Affected
387 devices (93 US)
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Brazil, Brunei Darussalam, Canada, China, Colombia, Croatia, Ecuador, Guatemala, India, Israel, Italy, Japan, Kazakhstan, Korea, Mexico, Morocco, New Zealand, Russian Federation, Spain, Taiwan, Thailand, United Kingdom, and Viet Nam
Location
Brooklyn Park, MN

Frequently Asked Questions

What product was recalled?
Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console, Model Numbers: 560BC, 560BC1, 560BCS, 560BCS1, R560BCS1. Used in cardiopulmonary bypass (CPB) procedures.. Recalled by Medtronic Perfusion Systems. Units affected: 387 devices (93 US).
Why was this product recalled?
An electrical component within the system controller module of certain Bio-Console 560 devices may exhibit electrical failure causing the device to stop pumping. Issues observed include the pump stopping, the user interface going blank, and smoking and producing a burning odor during use.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 7, 2021. Severity: Critical. Recall number: Z-1313-2021.