Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mL

Reported: March 27, 2024 Initiated: January 9, 2024 #Z-1314-2024

Product Description

Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mL

Reason for Recall

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Details

Units Affected
1 unit
Distribution
US Nationwide distribution in the states of AZ, PA and TX.
Location
Richmond, VA

Frequently Asked Questions

What product was recalled?
Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mL. Recalled by Mckesson Medical-Surgical Inc. Corporate Office. Units affected: 1 unit.
Why was this product recalled?
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 27, 2024. Severity: Moderate. Recall number: Z-1314-2024.