Severity
Moderate
FDA Devices recall · Reported February 18, 2026
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued …
Philips North America Llc recalled IQon Spectral CT; Product Code (REF): 728332; - a moderate-severity action.
IQon Spectral CT; Product Code (REF): 728332; was recalled by Philips North America Llc in February 18, 2026. Reason: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing…. Check the official notice for the remedy. Verify recall #Z-1316-2026 with the FDA Devices before acting.
The recall
Philips North America Llc issued this moderate-severity FDA Devices recall-As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1316-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1316-2026) was formally reported on February 18, 2026, with the manufacturer initiating the action on January 7, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips North America Llc is listed as the recalling firm, operating out of Cambridge, MA. Federal records list the affected scope as 52 units.
The documented reason for this recall is: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners … Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, B…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
52 units
Related Recalls
6
3 from same agency
IQon Spectral CT; Product Code (REF): 728332;
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1316-2026 |
| Date reported | February 18, 2026 |
| Date initiated | January 7, 2026 |
| Recalling firm | Philips North America Llc |
| Firm location | Cambridge, MA |
| Affected scope | 52 units |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, Chin… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 18, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.