Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a…
3M Company - Health Care Business recalled 3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas … - a moderate-severity action.
3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas … was recalled by 3M Company - Health Care Business in April 9, 2014. Reason: Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installati…. Check the official notice for the remedy. Verify recall #Z-1318-2014 with the FDA Devices before acting.
The recall
3M Company - Health Care Business issued this moderate-severity FDA Devices recall-Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installati….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1318-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1318-2014) was formally reported on April 9, 2014, with the manufacturer initiating the action on March 17, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. 3M Company - Health Care Business is listed as the recalling firm, operating out of Saint Paul, MN. Federal records list the affected scope as 73 US, 346 OUS.
The documented reason for this recall is: Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of t… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI. and the countries …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
73 US, 346 OUS
Related Recalls
6
3 from same agency
3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for INDOOR USE ONLY.
Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly. This assembly was only in production during the period November 10, 2006 through December 14, 2007 and in service during the period February 7, 20
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1318-2014 |
| Date reported | April 9, 2014 |
| Date initiated | March 17, 2014 |
| Recalling firm | 3M Company - Health Care Business |
| Firm location | Saint Paul, MN |
| Affected scope | 73 US, 346 OUS |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI. and the countries of COLOMBIA, VENEZUE… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 9, 2014. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.