Progressa Bed - AC-powered adjustable hospital bed
Reported: April 18, 2018 Initiated: January 12, 2018 #Z-1319-2018
Product Description
Progressa Bed - AC-powered adjustable hospital bed
Reason for Recall
A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.
Details
- Recalling Firm
- Hill-Rom, Inc.
- Units Affected
- 2117
- Distribution
- Worldwide distribution. US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.
- Location
- Batesville, IN
Frequently Asked Questions
What product was recalled? ▼
Progressa Bed - AC-powered adjustable hospital bed. Recalled by Hill-Rom, Inc.. Units affected: 2117.
Why was this product recalled? ▼
A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on April 18, 2018. Severity: Moderate. Recall number: Z-1319-2018.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11