DigitalDiagnost 4 High Performance. radiography and fluoroscopy system

Reported: July 20, 2022 Initiated: May 19, 2021 #Z-1326-2022 140 systems in total, (Updated 1/30/2023). units

Philips Medical Systems DMC GmbH issued this FDA Devices recall on July 20, 2022. Classified as Moderate severity (Class II). Approximately 140 systems in total, (Updated 1/30/2023). units are affected. The recall was issued because: Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, op…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1326-2022) was formally reported on July 20, 2022, with the manufacturer initiating the action on May 19, 2021. It is classified under Moderate severity (Class II), with a current status of Ongoing. Philips Medical Systems DMC GmbH is listed as the recalling firm, operating out of Hamburg. Federal records indicate 140 systems in total, (Updated 1/30/2023). units are affected, placing this recall in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps.

The documented reason for this recall is: Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the f… Distribution data in the federal record shows the product reached: US Nationwide Distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 4 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

140 systems in total, (Updated 1/30/2023).

Related Recalls

6

6 from same agency

Product Description

DigitalDiagnost 4 High Performance. radiography and fluoroscopy system

Reason for Recall

Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.

Details

Units Affected
140 systems in total, (Updated 1/30/2023).
Distribution
US Nationwide Distribution
Location
Hamburg

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1326-2022
Date reported July 20, 2022
Date initiated May 19, 2021
Recalling firm Philips Medical Systems DMC GmbH
Units affected 140 systems in total, (Updated 1/30/2023).
Distribution US Nationwide Distribution

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

140 systems in total, (Updated 1/30/2023). units affected — million-unit bracket.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units) ✓ This recall

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
DigitalDiagnost 4 High Performance. radiography and fluoroscopy system. Recalled by Philips Medical Systems DMC GmbH. Units affected: 140 systems in total, (Updated 1/30/2023)..
Why was this product recalled?
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 20, 2022. Severity: Moderate. Recall number: Z-1326-2022.
Where was the recalled product distributed?
Distribution: US Nationwide Distribution.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1326-2022) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).