Severity
Moderate
FDA Devices recall · Reported April 20, 2016
Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of th…
Siemens Medical Solutions USA, Inc recalled SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images … - a moderate-severity action.
SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images … was recalled by Siemens Medical Solutions USA, Inc in April 20, 2016. Reason: Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible …. Check the official notice for the remedy. Verify recall #Z-1350-2016 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc issued this moderate-severity FDA Devices recall-Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1350-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1350-2016) was formally reported on April 20, 2016, with the manufacturer initiating the action on March 22, 2016. It is classified under Moderate severity (Class II), with a current status of Terminated. Siemens Medical Solutions USA, Inc is listed as the recalling firm, operating out of Malvern, PA. Federal records list the affected scope as 1 CT system.
The documented reason for this recall is: Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened dur… Distribution data in the federal record shows the product reached: Distributed to: Cincinnati, OH.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1 CT system
Related Recalls
6
3 from same agency
SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from either the same axial plane taken at different angles or spiral planes taken at different angles.
Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1350-2016 |
| Date reported | April 20, 2016 |
| Date initiated | March 22, 2016 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Firm location | Malvern, PA |
| Affected scope | 1 CT system |
| Distribution | Distributed to: Cincinnati, OH. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 20, 2016. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.