Severity
Critical
FDA Devices recall · Reported March 11, 2020
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery ala…
CareFusion 303, Inc. recalled CareFusion Alaris Pump Module, Model 8100 - a critical-severity action.
CareFusion Alaris Pump Module, Model 8100 was recalled by CareFusion 303, Inc. in March 11, 2020. Reason: 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rat…. Check the official notice for the remedy. Verify recall #Z-1360-2020 with the FDA Devices before acting.
The recall
CareFusion 303, Inc. issued this critical-severity FDA Devices recall-1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rat….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1360-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1360-2020) was formally reported on March 11, 2020, with the manufacturer initiating the action on February 4, 2020. It is classified under Critical severity (Class I), with a current status of Ongoing. CareFusion 303, Inc. is listed as the recalling firm, operating out of San Diego, CA. Federal records list the affected scope as 1286898, placing this recall in the million-unit bracket that typically triggers nationwide consumer alerts and retailer sweeps.
The documented reason for this recall is: 1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority K… Distribution data in the federal record shows the product reached: Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
1286898
Related Recalls
6
3 from same agency
CareFusion Alaris Pump Module, Model 8100
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-1360-2020 |
| Date reported | March 11, 2020 |
| Date initiated | February 4, 2020 |
| Recalling firm | CareFusion 303, Inc. |
| Firm location | San Diego, CA |
| Affected scope | 1286898 |
| Distribution | Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 11, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.