Severity
Moderate
FDA Devices recall · Reported February 25, 2026
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, cra…
Medline Industries, LP recalled Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEED… - a moderate-severity action.
Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEED… was recalled by Medline Industries, LP in February 25, 2026. Reason: Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connector…. Check the official notice for the remedy. Verify recall #Z-1366-2026 with the FDA Devices before acting.
The recall
Medline Industries, LP issued this moderate-severity FDA Devices recall-Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connector….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1366-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1366-2026) was formally reported on February 25, 2026, with the manufacturer initiating the action on December 31, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Medline Industries, LP is listed as the recalling firm, operating out of Northfield, IL. Federal records list the affected scope as 926,227 units, a scale large enough to require multi-state distribution tracking.
The documented reason for this recall is: Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may … Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the country of Bahamas.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices and 2 from Medline Industries, LP - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
926,227 units
Related Recalls
6
3 from same agency
Medline Industries IV Administration and Extension Sets: 1) DBD-IV ADMIN SET,60DRP,1NEEDLE-FREE VALV, Model Number: DYNDTB0529; 2) IV ADMIN SET, 10DRP, 1NF, 92, Model Number: DYNDTB1029; 3) DBD-IVEXTSET2.4ML2 NEEDLE-FREEVLVE,35, Model Number: DYNDTB1516; 4) DBD-IV ADMIN SET, 15DRP, 1 INJECT PORT,, Model Number: DYNDTB1530; 5) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 6, Model Number: DYNDTB1539; 6) DBD-IV ADMIN SET, 15DRP, 3 NF, 109, Model Number: DYNDTB1545; 7) IV ADMIN SET, 20DRP, 3 NF, 109, Model Number: DYNDTB2045; 8) DBD-IVADMINSET60DRP2NEEDLE-FREEVLVE96, Model Number: DYNDTB0512; 9) DBD-IVADMINSET,60DRP3 NEEDLE-FREE VLVE11, Model Number: DYNDTB0555; 10) IV ADMIN SET, 10DRP, NF,INJ, 96, Model Number: DYNDTB1022; 11) DBD-IV ADMIN SET, 15DRP, 2NF, 96, Model Number: DYNDTB1512; 12) DBD-IV ADMIN SET, 15DRP, 1NF, 92, Model Number: DYNDTB1529; 13) DBD-IV ADMIN SET, 15DRP, ROLLER CLAMP, 9, Model Number: DYNDTB1537; 14) DBD-IV ADMIN SET, 15DRP, SECOND, 40, Model Number: DYNDTB1540; 15) IV ADMIN SET, 20DRP, 1NF, 92, Model Number: DYNDTB2029; 16) DBD-IV EXT SET, 6.5, 1.1ML, M LUERLOCK,, Model Number: DYNDTB5077; 17) IV ADMIN SET, 10DRP, 3 NF, 109, Model Number: DYNDTB1045; 18) DBD-IVADMISET15DRP4NEEDLEFREE VLVESTPCK1, Model Number: DYNDTB1545D; 19) IV ADMIN SET15DRPFLWCTRL1NEEDLEFREE VA, Model Number: DYNDTB2524; 20) IV ADMIN SET, 10DRP, 2NF, 96, Model Number: DYNDTB1012; 21) IV ADMIN SET, 20DRP, 2NF, 96, Model Number: DYNDTB2012; 22) IV ADMIN SET, 20DRP, NF,INJ, 96, Model Number: DYNDTB2022; 23) DBD-IV EXT SET,SMBORE EXT,8,M LUERLOCK,1, Model Number: DYNDTB5081
Under circumstances involving exposure to particular chemicals and mechanical forces, the male luer connectors for the specified IV Administration and Extension sets may leak, crack, and/or break during use, which may result in a delay in patient treatment, blood loss, infection, an air embolism, and/or underdosage of medication.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1366-2026 |
| Date reported | February 25, 2026 |
| Date initiated | December 31, 2025 |
| Recalling firm | Medline Industries, LP |
| Firm location | Northfield, IL |
| Affected scope | 926,227 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Bahamas. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported February 25, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.