Severity
Moderate
FDA Devices recall · Reported May 29, 2013
When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific …
Roche Diagnostics Operations, Inc. recalled Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the … - a moderate-severity action.
Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the … was recalled by Roche Diagnostics Operations, Inc. in May 29, 2013. Reason: When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead …. Check the official notice for the remedy. Verify recall #Z-1368-2013 with the FDA Devices before acting.
The recall
Roche Diagnostics Operations, Inc. issued this moderate-severity FDA Devices recall-When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1368-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1368-2013) was formally reported on May 29, 2013, with the manufacturer initiating the action on April 4, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Operations, Inc. is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 72 units.
The documented reason for this recall is: When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following co… Distribution data in the federal record shows the product reached: USA Nationwide Distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
72 units
Related Recalls
6
3 from same agency
Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes and photometric analysis modules
When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1368-2013 |
| Date reported | May 29, 2013 |
| Date initiated | April 4, 2013 |
| Recalling firm | Roche Diagnostics Operations, Inc. |
| Firm location | Indianapolis, IN |
| Affected scope | 72 units |
| Distribution | USA Nationwide Distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 29, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.