FDA Devices recall · Reported May 29, 2019
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when usi…
Roche Diagnostics Corporation recalled Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Dat… - a moderate-severity action.
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Dat… was recalled by Roche Diagnostics Corporation in May 29, 2019. Reason: Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the re…. Check the official notice for the remedy. Verify recall #Z-1368-2019 with the FDA Devices before acting.
The recall
Roche Diagnostics Corporation issued this moderate-severity FDA Devices recall-Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the re….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1368-2019 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1368-2019) was formally reported on May 29, 2019, with the manufacturer initiating the action on February 4, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Roche Diagnostics Corporation is listed as the recalling firm, operating out of Indianapolis, IN. Federal records list the affected scope as 15 units.
The documented reason for this recall is: Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does no… Distribution data in the federal record shows the product reached: US Distribution to states of: AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN and TX.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
15 units
Related Recalls
6
6 from same agency
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1368-2019 |
| Date reported | May 29, 2019 |
| Date initiated | February 4, 2019 |
| Recalling firm | Roche Diagnostics Corporation |
| Firm location | Indianapolis, IN |
| Affected scope | 15 units |
| Distribution | US Distribution to states of: AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN and TX. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 29, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.