Severity
Moderate
FDA Devices recall · Reported April 14, 2021
Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the…
Becton Dickinson & Co. recalled BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic N… - a moderate-severity action.
BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic N… was recalled by Becton Dickinson & Co. in April 14, 2021. Reason: Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reacti…. Check the official notice for the remedy. Verify recall #Z-1377-2021 with the FDA Devices before acting.
The recall
Becton Dickinson & Co. issued this moderate-severity FDA Devices recall-Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reacti….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1377-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1377-2021) was formally reported on April 14, 2021, with the manufacturer initiating the action on March 17, 2021. It is classified under Moderate severity (Class II), with a current status of Terminated. Becton Dickinson & Co. is listed as the recalling firm, operating out of Sparks, MD. Federal records list the affected scope as 15550 EA.
The documented reason for this recall is: Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributin… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
15550 EA
Related Recalls
6
3 from same agency
BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221558
Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1377-2021 |
| Date reported | April 14, 2021 |
| Date initiated | March 17, 2021 |
| Recalling firm | Becton Dickinson & Co. |
| Firm location | Sparks, MD |
| Affected scope | 15550 EA |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 14, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.