Severity
Moderate
FDA Devices recall · Reported April 25, 2018
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex …
Cook Inc. recalled MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocesso… - a moderate-severity action.
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocesso… was recalled by Cook Inc. in April 25, 2018. Reason: The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for pr…. Check the official notice for the remedy. Verify recall #Z-1378-2018 with the FDA Devices before acting.
The recall
Cook Inc. issued this moderate-severity FDA Devices recall-The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for pr….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1378-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1378-2018) was formally reported on April 25, 2018, with the manufacturer initiating the action on January 12, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Cook Inc. is listed as the recalling firm, operating out of Bloomington, IN. Federal records list the affected scope as 4,451.
The documented reason for this recall is: The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater. Distribution data in the federal record shows the product reached: Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide). Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4,451
Related Recalls
6
3 from same agency
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1378-2018 |
| Date reported | April 25, 2018 |
| Date initiated | January 12, 2018 |
| Recalling firm | Cook Inc. |
| Firm location | Bloomington, IN |
| Affected scope | 4,451 |
| Distribution | Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide) |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 25, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.