Severity
Moderate
FDA Devices recall · Reported April 25, 2018
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM…
Ge Medical Systems Ultrasound & Primary Care Diagnostics, Ll recalled GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DP… - a moderate-severity action.
GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DP… was recalled by Ge Medical Systems Ultrasound & Primary Care Diagnostics, Ll in April 25, 2018. Reason: Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam m…. Check the official notice for the remedy. Verify recall #Z-1394-2018 with the FDA Devices before acting.
The recall
Ge Medical Systems Ultrasound & Primary Care Diagnostics, Ll issued this moderate-severity FDA Devices recall-Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam m….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1394-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1394-2018) was formally reported on April 25, 2018, with the manufacturer initiating the action on March 16, 2018. It is classified under Moderate severity (Class II), with a current status of Terminated. Ge Medical Systems Ultrasound & Primary Care Diagnostics, Ll is listed as the recalling firm, operating out of Madison, WI. Federal records list the affected scope as 43 units.
The documented reason for this recall is: Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kon…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
43 units
Related Recalls
6
3 from same agency
GE Healthcare Lunar: a) DPX NT, Model Numbers: LU8230, LU40338, LU42357, LU42369 b) DPX MD+, Model Numbers: LU8230, LU40338, LU40352 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1394-2018 |
| Date reported | April 25, 2018 |
| Date initiated | March 16, 2018 |
| Recalling firm | Ge Medical Systems Ultrasound & Primary Care Diagnostics, Ll |
| Firm location | Madison, WI |
| Affected scope | 43 units |
| Distribution | Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, I… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 25, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.