FDA Devices recall · Reported March 15, 2017
AngioJet¿ Ultra 5000 A Console:Foot Switch Material number (U PN) 105650 Cardiovascular: The Console is a multiple-use device that controls the Thrombectomy Set. It drives the pump, regulates fluid inflow and outflow, provides the operator with Ultra System set-up prompts, total infused saline volume, and Ultra System malfunction warnings.
Boston Scientific has received field reports regarding the AngioJet Ultra 5000 A Console foot switch sticking during use resulting in an undesired continued operation of the devic…
Boston Scientific Corporation recalled AngioJet¿ Ultra 5000 A Console:Foot Switch Material number (U PN) 105650 Cardiovascul… - a moderate-severity action.
AngioJet¿ Ultra 5000 A Console:Foot Switch Material number (U PN) 105650 Cardiovascul… was recalled by Boston Scientific Corporation in March 15, 2017. Reason: Boston Scientific has received field reports regarding the AngioJet Ultra 5000 A Console foot switch sticking…. Check the official notice for the remedy. Verify recall #Z-1401-2017 with the FDA Devices before acting.
The recall
Boston Scientific Corporation issued this moderate-severity FDA Devices recall-Boston Scientific has received field reports regarding the AngioJet Ultra 5000 A Console foot switch sticking….
- Moderate
- severity level
- 3K units
- affected scope
- Class II
- classification
- March 15, 2017
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1401-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
AngioJet¿ Ultra 5000 A Console:Foot Switch Material number (U PN) 105650 Cardiovascular: The Console is a multiple-use device that controls the Thrombectomy Set. It drives the pump, regulates fluid inflow and outflow, provides the operator with Ultra System set-up prompts, total infused saline volume, and Ultra System malfunction warnings.
Reason for recall
Boston Scientific has received field reports regarding the AngioJet Ultra 5000 A Console foot switch sticking during use resulting in an undesired continued operation of the device.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1401-2017 |
| Date reported | March 15, 2017 |
| Date initiated | January 31, 2017 |
| Recalling firm | Boston Scientific Corporation |
| Firm location | Fremont, CA |
| Affected scope | 3402 units; 2626 in US; 786 units OUS |
| Distribution | US and Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Denmark, Dominican Rep., Ecuador, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Ind… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1401-2017) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026