Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Reported: April 21, 2021 Initiated: March 5, 2021 #Z-1411-2021

Product Description

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Reason for Recall

Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

Details

Recalling Firm
Covidien
Units Affected
59,828 devices
Distribution
Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
Location
Ponce, PR

Frequently Asked Questions

What product was recalled?
Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094. Recalled by Covidien. Units affected: 59,828 devices.
Why was this product recalled?
Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 21, 2021. Severity: Moderate. Recall number: Z-1411-2021.