namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Recall #
Z-1431-2025
Affected scope
1669 units
Initiated
February 12, 2025
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Medline Industries, LP - Northfield recalled namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468;… - a moderate-severity action.

namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468;… was recalled by Medline Industries, LP - Northfield in April 2, 2025. Reason: The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near …. Check the official notice for the remedy. Verify recall #Z-1431-2025 with the FDA Devices before acting.

The recall

Medline Industries, LP - Northfield issued this moderate-severity FDA Devices recall-The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near ….

Moderate
severity level
2K units
affected scope
Class II
classification
April 2, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1431-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1431-2025) was formally reported on April 2, 2025, with the manufacturer initiating the action on February 12, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Medline Industries, LP - Northfield is listed as the recalling firm, operating out of Northfield, IL. Federal records list the affected scope as 1669 units.

The documented reason for this recall is: The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway. Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

1669 units

Related Recalls

6

3 from same agency

Product description

namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 11) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 12) MTS,LEFT HEART,MAN,KIT,S*-BALTIMORE, REF 60190947; 13) MTS-LEFT HEART,KIT-W*,H* HOSP PG, REF 60231052; 14) MTO,3-VALVE,KIT-T*,A*-PG, REF 65190182; 15) MTO,LEFT HEART,KIT,-,S* F*,-,PG, REF 65190474; 16) MTS,DR.,MCKAYS,4V,KIT,-,H* HOSP PG, REF 600800516; 17) MTS,EP,KIT,-,S*,TEACHING HOSPITAL PG, REF 601309013; 18) MTS,RHT,HRT-S* HOSP &,MC, REF 601905413; 19) MTS,PEDS,LFT,&,RT,HRT-S* HOSP, REF 601905414; 20) MTO,SCH,BEL,J*,H*-ST-PAUL,-,PG, REF 606604328; 21) LHK,-,MID,3VCOMP,OFF,PS,S10RA, REF 613000122; 22) MTS,LHK,W* HOSPITAL, REF 6010034281; 23) MTS,LHK,W* HOSPITAL, REF 6010034281; 24) MTS,LHK,P*,P* MED CTR, REF 6010171931; 25) MTS,LHK,M* HEALTHCARE,MGMT, REF 6061073001; 26) TVR US B* INTEGRATED SERVICE CENT, REF 601316711A; 27) LHK-S* MEDICAL CENTER,A*, REF 60190776A; 28) LHK,-,S*,AND,W* HOSPITAL, REF 60191118A.

Reason for recall

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1431-2025
Date reported April 2, 2025
Date initiated February 12, 2025
Recalling firm Medline Industries, LP - Northfield
Firm location Northfield, IL
Affected scope 1669 units
Distribution Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1669 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1431-2025) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 11) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 12) MTS,LEFT HEART,MAN,KIT,S*-BALTIMORE, REF 60190947; 13) MTS-LEFT HEART,KIT-W*,H* HOSP PG, REF 60231052; 14) MTO,3-VALVE,KIT-T*,A*-PG, REF 65190182; 15) MTO,LEFT HEART,KIT,-,S* F*,-,PG, REF 65190474; 16) MTS,DR.,MCKAYS,4V,KIT,-,H* HOSP PG, REF 600800516; 17) MTS,EP,KIT,-,S*,TEACHING HOSPITAL PG, REF 601309013; 18) MTS,RHT,HRT-S* HOSP &,MC, REF 601905413; 19) MTS,PEDS,LFT,&,RT,HRT-S* HOSP, REF 601905414; 20) MTO,SCH,BEL,J*,H*-ST-PAUL,-,PG, REF 606604328; 21) LHK,-,MID,3VCOMP,OFF,PS,S10RA, REF 613000122; 22) MTS,LHK,W* HOSPITAL, REF 6010034281; 23) MTS,LHK,W* HOSPITAL, REF 6010034281; 24) MTS,LHK,P*,P* MED CTR, REF 6010171931; 25) MTS,LHK,M* HEALTHCARE,MGMT, REF 6061073001; 26) TVR US B* INTEGRATED SERVICE CENT, REF 601316711A; 27) LHK-S* MEDICAL CENTER,A*, REF 60190776A; 28) LHK,-,S*,AND,W* HOSPITAL, REF 60191118A.. Recalled by Medline Industries, LP - Northfield. Units affected: 1669 units.
Why was this product recalled?
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 2, 2025. Severity: Moderate. Recall number: Z-1431-2025.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1431-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 2, 2025. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.