namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT

Reported: April 2, 2025 Initiated: February 12, 2025 #Z-1431-2025 1669 units units

MEDLINE INDUSTRIES, LP - Northfield issued this FDA Devices recall on April 2, 2025. Classified as Moderate severity (Class II). Approximately 1669 units units are affected. The recall was issued because: The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid …. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1431-2025) was formally reported on April 2, 2025, with the manufacturer initiating the action on February 12, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. MEDLINE INDUSTRIES, LP - Northfield is listed as the recalling firm, operating out of Northfield, IL. Federal records indicate 1669 units units are affected.

The documented reason for this recall is: The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway. Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 1 year old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

1669 units

Related Recalls

6

6 from same agency

Product Description

namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 11) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 12) MTS,LEFT HEART,MAN,KIT,S*-BALTIMORE, REF 60190947; 13) MTS-LEFT HEART,KIT-W*,H* HOSP PG, REF 60231052; 14) MTO,3-VALVE,KIT-T*,A*-PG, REF 65190182; 15) MTO,LEFT HEART,KIT,-,S* F*,-,PG, REF 65190474; 16) MTS,DR.,MCKAYS,4V,KIT,-,H* HOSP PG, REF 600800516; 17) MTS,EP,KIT,-,S*,TEACHING HOSPITAL PG, REF 601309013; 18) MTS,RHT,HRT-S* HOSP &,MC, REF 601905413; 19) MTS,PEDS,LFT,&,RT,HRT-S* HOSP, REF 601905414; 20) MTO,SCH,BEL,J*,H*-ST-PAUL,-,PG, REF 606604328; 21) LHK,-,MID,3VCOMP,OFF,PS,S10RA, REF 613000122; 22) MTS,LHK,W* HOSPITAL, REF 6010034281; 23) MTS,LHK,W* HOSPITAL, REF 6010034281; 24) MTS,LHK,P*,P* MED CTR, REF 6010171931; 25) MTS,LHK,M* HEALTHCARE,MGMT, REF 6061073001; 26) TVR US B* INTEGRATED SERVICE CENT, REF 601316711A; 27) LHK-S* MEDICAL CENTER,A*, REF 60190776A; 28) LHK,-,S*,AND,W* HOSPITAL, REF 60191118A.

Reason for Recall

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

Details

Units Affected
1669 units
Distribution
Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
Location
Northfield, IL

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1431-2025
Date reported April 2, 2025
Date initiated February 12, 2025
Recalling firm MEDLINE INDUSTRIES, LP - Northfield
Units affected 1669 units
Distribution Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1669 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 9) MTS,3V,KIT-M*,MEMORIAL, REF 60131208; 10) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 11) MTO,LEFT HEART,KIT,-,N* HOSPITAL PG, REF 60140101; 12) MTS,LEFT HEART,MAN,KIT,S*-BALTIMORE, REF 60190947; 13) MTS-LEFT HEART,KIT-W*,H* HOSP PG, REF 60231052; 14) MTO,3-VALVE,KIT-T*,A*-PG, REF 65190182; 15) MTO,LEFT HEART,KIT,-,S* F*,-,PG, REF 65190474; 16) MTS,DR.,MCKAYS,4V,KIT,-,H* HOSP PG, REF 600800516; 17) MTS,EP,KIT,-,S*,TEACHING HOSPITAL PG, REF 601309013; 18) MTS,RHT,HRT-S* HOSP &,MC, REF 601905413; 19) MTS,PEDS,LFT,&,RT,HRT-S* HOSP, REF 601905414; 20) MTO,SCH,BEL,J*,H*-ST-PAUL,-,PG, REF 606604328; 21) LHK,-,MID,3VCOMP,OFF,PS,S10RA, REF 613000122; 22) MTS,LHK,W* HOSPITAL, REF 6010034281; 23) MTS,LHK,W* HOSPITAL, REF 6010034281; 24) MTS,LHK,P*,P* MED CTR, REF 6010171931; 25) MTS,LHK,M* HEALTHCARE,MGMT, REF 6061073001; 26) TVR US B* INTEGRATED SERVICE CENT, REF 601316711A; 27) LHK-S* MEDICAL CENTER,A*, REF 60190776A; 28) LHK,-,S*,AND,W* HOSPITAL, REF 60191118A.. Recalled by MEDLINE INDUSTRIES, LP - Northfield. Units affected: 1669 units.
Why was this product recalled?
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 2, 2025. Severity: Moderate. Recall number: Z-1431-2025.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1431-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).