FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.

Reported: April 16, 2014 Initiated: March 5, 2014 #Z-1441-2014

Product Description

FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.

Reason for Recall

Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).

Details

Recalling Firm
Remington Medical Inc.
Units Affected
41,499 cables
Distribution
USA (nationwide) Distribution including Canada
Location
Alpharetta, GA

Frequently Asked Questions

What product was recalled?
FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.. Recalled by Remington Medical Inc.. Units affected: 41,499 cables.
Why was this product recalled?
Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).
Which agency issued this recall?
This recall was issued by the FDA Devices on April 16, 2014. Severity: Moderate. Recall number: Z-1441-2014.