Severity
Moderate
FDA Devices recall · Reported June 12, 2013
Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This …
Intuitive Surgical, Inc. recalled 8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: I… - a moderate-severity action.
8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: I… was recalled by Intuitive Surgical, Inc. in June 12, 2013. Reason: Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and …. Check the official notice for the remedy. Verify recall #Z-1442-2013 with the FDA Devices before acting.
The recall
Intuitive Surgical, Inc. issued this moderate-severity FDA Devices recall-Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1442-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1442-2013) was formally reported on June 12, 2013, with the manufacturer initiating the action on May 8, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Intuitive Surgical, Inc. is listed as the recalling firm, operating out of Sunnyvale, CA. Federal records list the affected scope as ~132,441 instruments.
The documented reason for this recall is: Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical… Distribution data in the federal record shows the product reached: Worldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
~132,441 instruments
Related Recalls
6
3 from same agency
8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.
Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1442-2013 |
| Date reported | June 12, 2013 |
| Date initiated | May 8, 2013 |
| Recalling firm | Intuitive Surgical, Inc. |
| Firm location | Sunnyvale, CA |
| Affected scope | ~132,441 instruments |
| Distribution | Worldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt, Guadeloupe, Israel… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported June 12, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.