PlainRecalls
FDA Devices Moderate Class II Terminated

Hardy Diagnostics cat no. G327 HardyCHROM SS Agar (HardyCHROM Salmonella Shigella) Cat no: G327 Container type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2 to 8 degrees C. on receipt in dark A selective chromogenic medium for the primary screening isolation, and differentiation of Salmonella & Shigella spp.

Reported: April 22, 2015 Initiated: March 4, 2015 #Z-1444-2015

Product Description

Hardy Diagnostics cat no. G327 HardyCHROM SS Agar (HardyCHROM Salmonella Shigella) Cat no: G327 Container type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2 to 8 degrees C. on receipt in dark A selective chromogenic medium for the primary screening isolation, and differentiation of Salmonella & Shigella spp.

Reason for Recall

The firm is recalling HardyCHROM SS, Cat. no. G327 Lot no. 15037 due to a report of possible lack of color development in Shigella organisms in the referenced lot.

Details

Recalling Firm
Hardy Diagnostics
Units Affected
840 (84 pk/10)
Distribution
Worldwide Distribution -- U.S., including the states of TX, CA, AZ, WA, PR, and IA; and the countries of Canada and NSW.
Location
Santa Maria, CA

Frequently Asked Questions

What product was recalled?
Hardy Diagnostics cat no. G327 HardyCHROM SS Agar (HardyCHROM Salmonella Shigella) Cat no: G327 Container type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2 to 8 degrees C. on receipt in dark A selective chromogenic medium for the primary screening isolation, and differentiation of Salmonella & Shigella spp.. Recalled by Hardy Diagnostics. Units affected: 840 (84 pk/10).
Why was this product recalled?
The firm is recalling HardyCHROM SS, Cat. no. G327 Lot no. 15037 due to a report of possible lack of color development in Shigella organisms in the referenced lot.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 22, 2015. Severity: Moderate. Recall number: Z-1444-2015.