Severity
Moderate
FDA Devices recall · Reported April 29, 2015
Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out o…
Del Mar Reynolds Medical, Ltd. recalled Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcar… - a moderate-severity action.
Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcar… was recalled by Del Mar Reynolds Medical, Ltd. in April 29, 2015. Reason: Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume…. Check the official notice for the remedy. Verify recall #Z-1458-2015 with the FDA Devices before acting.
The recall
Del Mar Reynolds Medical, Ltd. issued this moderate-severity FDA Devices recall-Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1458-2015 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1458-2015) was formally reported on April 29, 2015, with the manufacturer initiating the action on March 5, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Del Mar Reynolds Medical, Ltd. is listed as the recalling firm, operating out of Hertford. Federal records list the affected scope as 1040 units total (398 in the US and 642 international).
The documented reason for this recall is: Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actu… Distribution data in the federal record shows the product reached: Worldwide Distribution-US including the states of Louisiana, North Carolina, and Washington and the countries of Argentina, Brazil, China, Colombia, Ecuador, Finland, Great Britain, India, Libya, Mexico, Nicaragua, Oman…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1040 units total (398 in the US and 642 international)
Related Recalls
6
3 from same agency
Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia.
Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1458-2015 |
| Date reported | April 29, 2015 |
| Date initiated | March 5, 2015 |
| Recalling firm | Del Mar Reynolds Medical, Ltd. |
| Firm location | Hertford |
| Affected scope | 1040 units total (398 in the US and 642 international) |
| Distribution | Worldwide Distribution-US including the states of Louisiana, North Carolina, and Washington and the countries of Argentina, Brazil, China, Colombia, Ecuador, Finland, Great Britain, India, Libya, Mexico, Nicaragua, Oman, Philippines, Saudi… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 29, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.