PlainRecalls
FDA Devices Moderate Class II Ongoing

AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

Reported: August 3, 2022 Initiated: June 16, 2022 #Z-1463-2022

Product Description

AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US

Reason for Recall

Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock

Details

Units Affected
277 units
Distribution
US Nationwide distribution.
Location
Waltham, MA

Frequently Asked Questions

What product was recalled?
AquaC UNO H [120V, 60Hz, 1700VA, Fresenius US-Version: 24-0970-1 Vivonic Part Number: G02000966-US. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 277 units.
Why was this product recalled?
Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock
Which agency issued this recall?
This recall was issued by the FDA Devices on August 3, 2022. Severity: Moderate. Recall number: Z-1463-2022.