Severity
Moderate
FDA Devices recall · Reported March 4, 2026
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result…
Medline Industries, LP recalled Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBS… - a moderate-severity action.
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBS… was recalled by Medline Industries, LP in March 4, 2026. Reason: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include s…. Check the official notice for the remedy. Verify recall #Z-1467-2026 with the FDA Devices before acting.
The recall
Medline Industries, LP issued this moderate-severity FDA Devices recall-Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include s….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1467-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1467-2026) was formally reported on March 4, 2026, with the manufacturer initiating the action on January 8, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Medline Industries, LP is listed as the recalling firm, operating out of Northfield, IL. Federal records list the affected scope as 31,848.
The documented reason for this recall is: Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by… Distribution data in the federal record shows the product reached: US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices and 2 from Medline Industries, LP - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
31,848
Related Recalls
6
3 from same agency
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1467-2026 |
| Date reported | March 4, 2026 |
| Date initiated | January 8, 2026 |
| Recalling firm | Medline Industries, LP |
| Firm location | Northfield, IL |
| Affected scope | 31,848 |
| Distribution | US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 4, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.