VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Reported: March 18, 2020 Initiated: August 12, 2018 #Z-1469-2020

Product Description

VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Reason for Recall

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Details

Units Affected
462 kits
Distribution
Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.
Location
Birmingham

Frequently Asked Questions

What product was recalled?
VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.. Recalled by The Binding Site Group, Ltd.. Units affected: 462 kits.
Why was this product recalled?
A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.
Which agency issued this recall?
This recall was issued by the FDA Devices on March 18, 2020. Severity: Moderate. Recall number: Z-1469-2020.