Severity
Moderate
FDA Devices recall · Reported March 22, 2017
Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.
Ortho-Clinical Diagnostics recalled VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (1… - a moderate-severity action.
VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (1… was recalled by Ortho-Clinical Diagnostics in March 22, 2017. Reason: Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.. Check the official notice for the remedy. Verify recall #Z-1471-2017 with the FDA Devices before acting.
The recall
Ortho-Clinical Diagnostics issued this moderate-severity FDA Devices recall-Intermittent imprecision in results generated using two specific lots of VITROS Li Slides..
Sourced from official FDA Devices enforcement records. Verify recall #Z-1471-2017 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1471-2017) was formally reported on March 22, 2017, with the manufacturer initiating the action on February 1, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Ortho-Clinical Diagnostics is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as US: 94 sales units; OUS: 136 sales units.
The documented reason for this recall is: Intermittent imprecision in results generated using two specific lots of VITROS Li Slides. Distribution data in the federal record shows the product reached: US, Puerto Rico, Canada, EU, China, Brazil, Chile, and Dominican Republic.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
US: 94 sales units; OUS: 136 sales units
Related Recalls
6
3 from same agency
VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90 slides per sales unit) and REF/Catalog Number/Product Code 8318925 (60 slide format / 300 slides per sales unit), Unique Identifier Nos. 10758750009466 and 10758750004720, RX ONLY, IVD --Ortho Clinical Diagnostics, Inc. Rochester, NY 14626 --- Background: The VITROS Li Slide is a multilayered, analytical element coated on a polyester support. During the coating process, thin layers of reagents are coated onto a polyester support layer to create a spreading layer and a chemical (reagent) layer. During the coating process rolls of materials are produced. These rolls are segmented into slits. There are several slits in each roll. Each slit is then cut into small chemistry chips that are positioned between plastic to create the final Microslide format. Eighteen (18) or sixty (60) slides are placed into each plastic cartridge for use. Each slide is a device.
Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1471-2017 |
| Date reported | March 22, 2017 |
| Date initiated | February 1, 2017 |
| Recalling firm | Ortho-Clinical Diagnostics |
| Firm location | Rochester, NY |
| Affected scope | US: 94 sales units; OUS: 136 sales units |
| Distribution | US, Puerto Rico, Canada, EU, China, Brazil, Chile, and Dominican Republic. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 22, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.