FDA Devices recall · Reported March 11, 2026
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electrom…
Olympus Corporation of the Americas recalled Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockP… - a moderate-severity action.
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockP… was recalled by Olympus Corporation of the Americas in March 11, 2026. Reason: Mis-wired component-the improperly wired component can result in additional noise on the power supply output.…. Check the official notice for the remedy. Verify recall #Z-1494-2026 with the FDA Devices before acting.
The recall
Olympus Corporation of the Americas issued this moderate-severity FDA Devices recall-Mis-wired component-the improperly wired component can result in additional noise on the power supply output.….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1494-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1494-2026) was formally reported on March 11, 2026, with the manufacturer initiating the action on January 8, 2026. It is classified under Moderate severity (Class II), with a current status of Ongoing. Olympus Corporation of the Americas is listed as the recalling firm, operating out of Center Valley, PA. Federal records list the affected scope as 55 units (All OUS).
The documented reason for this recall is: Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards an… Distribution data in the federal record shows the product reached: International distribution in the country of Canada, Germany, Singapore, Australia, and India.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
55 units (All OUS)
Related Recalls
6
3 from same agency
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system.
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1494-2026 |
| Date reported | March 11, 2026 |
| Date initiated | January 8, 2026 |
| Recalling firm | Olympus Corporation of the Americas |
| Firm location | Center Valley, PA |
| Affected scope | 55 units (All OUS) |
| Distribution | International distribution in the country of Canada, Germany, Singapore, Australia, and India. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 11, 2026. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.