FDA Devices recall · Reported May 5, 2021
Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Model Numbers: System Material # AXIOM Artis FC 05904433; AXIOM Artis FA 05904441; AXIOM Artis MP 05904466; AXIOM Artis BC 05904649; AXIOM Artis BA 05904656; AXIOM Artis dBC Mag. Nav. 05917054; AXIOM Artis TA 07007755; AXIOM Artis dTA 07008605; AXIOM Artis dFC 07412807; AXIO
Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a …
Siemens Medical Solutions USA, Inc recalled Artis systems - Product Usage: an angiography systems developed for single and biplane di… - a moderate-severity action.
Artis systems - Product Usage: an angiography systems developed for single and biplane di… was recalled by Siemens Medical Solutions USA, Inc in May 5, 2021. Reason: Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed…. Check the official notice for the remedy. Verify recall #Z-1496-2021 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc issued this moderate-severity FDA Devices recall-Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed….
- Moderate
- severity level
- 3K units
- affected scope
- Class II
- classification
- May 5, 2021
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1496-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Model Numbers: System Material # AXIOM Artis FC 05904433; AXIOM Artis FA 05904441; AXIOM Artis MP 05904466; AXIOM Artis BC 05904649; AXIOM Artis BA 05904656; AXIOM Artis dBC Mag. Nav. 05917054; AXIOM Artis TA 07007755; AXIOM Artis dTA 07008605; AXIOM Artis dFC 07412807; AXIOM Artis dTC 07413078; AXIOM Artis dBA 07555357; AXIOM Artis dMP 07555365; AXIOM Artis dFA 07555373; AXIOM Artis dFC Mag. Nav. 07727717; AXIOM Artis TC 07728350; AXIOM Artis dBC 07728392; Artis zee floor 10094135; Artis zee ceiling 10094137; Artis zee MP 10094139; Artis zee biplane 10094141; Artis zee floor MN 10094142; Artis zee biplane MN 10094143; Artis Q floor 10848280; Artis Q ceiling 10848281; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen ceiling 10848354; Artis Q.zen biplane 10848355.
Reason for recall
Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited functionality of the Artis system up to system failure. Unintended movement of the C-Arm may cause a hazardous situation to the patient, operator, or staff members and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1496-2021 |
| Date reported | May 5, 2021 |
| Date initiated | March 30, 2021 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Firm location | Malvern, PA |
| Affected scope | 3,320 units U.S. |
| Distribution | US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1496-2021) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026