Severity
Moderate
FDA Devices recall · Reported March 25, 2020
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped…
Philips North America, LLC recalled HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC,… - a moderate-severity action.
HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC,… was recalled by Philips North America, LLC in March 25, 2020. Reason: Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal…. Check the official notice for the remedy. Verify recall #Z-1499-2020 with the FDA Devices before acting.
The recall
Philips North America, LLC issued this moderate-severity FDA Devices recall-Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1499-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1499-2020) was formally reported on March 25, 2020, with the manufacturer initiating the action on February 26, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips North America, LLC is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 97788 units.
The documented reason for this recall is: Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical sho… Distribution data in the federal record shows the product reached: US Nationwide domestic distribution. Worldwide foreign distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
97788 units
Related Recalls
6
3 from same agency
HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9.
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1499-2020 |
| Date reported | March 25, 2020 |
| Date initiated | February 26, 2020 |
| Recalling firm | Philips North America, LLC |
| Firm location | Andover, MA |
| Affected scope | 97788 units |
| Distribution | US Nationwide domestic distribution. Worldwide foreign distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 25, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.