BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897,
Product Description
BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784
Reason for Recall
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Details
- Recalling Firm
- Bard Access Systems, Inc.
- Units Affected
- 718,461
- Distribution
- Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- Location
- Salt Lake City, UT
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11