PlainRecalls

The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid

Reported: August 17, 2022 Initiated: June 24, 2022 #Z-1508-2022 436,990 units

Alere San Diego, Inc. issued this FDA Devices recall on August 17, 2022. Classified as Moderate severity (Class II). Approximately 436,990 units are affected. The recall was issued because: Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was f…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-1508-2022) was formally reported on August 17, 2022, with the manufacturer initiating the action on June 24, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Alere San Diego, Inc. is listed as the recalling firm, operating out of San Diego, CA. Federal records indicate 436,990 units are affected, a scale large enough to require multi-state distribution tracking.

The documented reason for this recall is: Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6… Distribution data in the federal record shows the product reached: US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 4 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

436,990

Related Recalls

6

6 from same agency

Product Description

The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987

Reason for Recall

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Details

Recalling Firm
Alere San Diego, Inc.
Units Affected
436,990
Distribution
US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA
Location
San Diego, CA

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-1508-2022
Date reported August 17, 2022
Date initiated June 24, 2022
Recalling firm Alere San Diego, Inc.
Units affected 436,990
Distribution US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR,…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

436,990 units affected — multi-state distribution scale.

Regional (<10K units)
Multi-state (10K – 100K units)
Large-scale (100K – 1M units) ✓ This recall
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987. Recalled by Alere San Diego, Inc.. Units affected: 436,990.
Why was this product recalled?
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Which agency issued this recall?
This recall was issued by the FDA Devices on August 17, 2022. Severity: Moderate. Recall number: Z-1508-2022.
Where was the recalled product distributed?
Distribution: US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1508-2022) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).