Severity
Moderate
Alere San Diego, Inc. issued this FDA Devices recall on August 17, 2022. Classified as Moderate severity (Class II). Approximately 436,990 units are affected. The recall was issued because: Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was f…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.
This FDA Devices action (record #Z-1508-2022) was formally reported on August 17, 2022, with the manufacturer initiating the action on June 24, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Alere San Diego, Inc. is listed as the recalling firm, operating out of San Diego, CA. Federal records indicate 436,990 units are affected, a scale large enough to require multi-state distribution tracking.
The documented reason for this recall is: Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6… Distribution data in the federal record shows the product reached: US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 4 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.
Severity
Moderate
Units Affected
436,990
Related Recalls
6
6 from same agency
The Cholestech LDX¿ System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX¿ cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1508-2022 |
| Date reported | August 17, 2022 |
| Date initiated | June 24, 2022 |
| Recalling firm | Alere San Diego, Inc. |
| Units affected | 436,990 |
| Distribution | US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
Learn how the US recall system works and how to protect yourself and your household.
How the US Recall System Works
The three federal agencies, how recalls are initiated, and what happens next.
Understanding Recall Severity Classes
What Class I, II, and III mean and which recalls demand immediate action.
What to Do When a Product Is Recalled
Verify, claim your remedy, report injuries, and navigate the process.
How to Check If Your Products Are Recalled
Step-by-step guide to checking food, products, medications, and vehicles.
Recalled Products in Your Home
A room-by-room household audit guide for active recalls.
Most Recalled Product Categories
Rankings of highest-recall categories from FDA, CPSC, and NHTSA.
Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.
Explore related product safety and public health data from federal sources.
Food Safety Inspections
FDA and state food facility inspection results — violations, enforcement actions, and compliance history for food manufacturers and processors.
PlainFoodSafe →
Ingredient Safety Data
FDA food additive and ingredient safety database — regulatory status, usage limits, and safety assessments for thousands of ingredients.
PlainIngredients →
Nutrition Data
USDA nutrition facts for 2M+ food items — calories, macros, vitamins, and ingredient analysis to verify what you consume.
GetFoodFacts →
Drug Safety Information
FDA drug data for 680+ medications — interactions, alternatives, side effects, and safety information for recalled and active drugs.
PlainMeds →
Product Injury Data
CPSC emergency room injury data for 838 product categories — 7.3M NEISS records tracking real-world consumer product injuries.
PlainSafety →
Recall Checker
Search our database of 83K+ recalls by product name, brand, or recall number across all agencies.
Check Now →
Recall Radar
Live feed of the latest recalls across FDA, CPSC, NHTSA, and USDA — filter by agency and severity.
View Radar →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
Waldemar Link GmbH & Co. KG (Mfg Site) · 2026-03-18
Penner Patient Care, Inc. · 2026-03-18
Instrumentation Laboratory · 2026-03-18
Beckman Coulter Mishima K.K. · 2026-03-18
Trividia Health, Inc. · 2026-03-18
Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →
Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Read our methodology — how this data is sourced, computed, and verified.
Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).