Severity
Moderate
FDA Devices recall · Reported March 25, 2020
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed thes…
Boston Scientific Corporation recalled Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Sci… - a moderate-severity action.
Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Sci… was recalled by Boston Scientific Corporation in March 25, 2020. Reason: Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Mode…. Check the official notice for the remedy. Verify recall #Z-1511-2020 with the FDA Devices before acting.
The recall
Boston Scientific Corporation issued this moderate-severity FDA Devices recall-Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Mode….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1511-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1511-2020) was formally reported on March 25, 2020, with the manufacturer initiating the action on February 12, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Boston Scientific Corporation is listed as the recalling firm, operating out of Saint Paul, MN. Federal records list the affected scope as 4 devices.
The documented reason for this recall is: Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software appli… Distribution data in the federal record shows the product reached: US Nationwide distribution in the states of NC, TN, and MO.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
4 devices
Related Recalls
6
3 from same agency
Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1511-2020 |
| Date reported | March 25, 2020 |
| Date initiated | February 12, 2020 |
| Recalling firm | Boston Scientific Corporation |
| Firm location | Saint Paul, MN |
| Affected scope | 4 devices |
| Distribution | US Nationwide distribution in the states of NC, TN, and MO. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported March 25, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.