VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY

Reported: May 7, 2014 Initiated: March 26, 2014 #Z-1519-2014

Product Description

VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY

Reason for Recall

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.

Details

Units Affected
1656 distributed (1200 went to warehouse in France)
Distribution
Nationwide, Puerto Rico, and foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY. Recalled by Ortho-Clinical Diagnostics. Units affected: 1656 distributed (1200 went to warehouse in France).
Why was this product recalled?
Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 7, 2014. Severity: Moderate. Recall number: Z-1519-2014.