Siemens SOMATOM Definition, SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
Reported: April 27, 2016 Initiated: March 2, 2016 #Z-1521-2016
Product Description
Siemens SOMATOM Definition, SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.
Reason for Recall
Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
Details
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Units Affected
- 1096
- Distribution
- Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
- Location
- Malvern, PA
Frequently Asked Questions
What product was recalled? ▼
Siemens SOMATOM Definition, SOMATOM Definition AS, SOMATOM Definition Flash and SOMATOM Definition Edge; Intended to produce cross-sectional images of the body. Computed tomography (CT) x-ray system.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 1096.
Why was this product recalled? ▼
Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on April 27, 2016. Severity: Moderate. Recall number: Z-1521-2016.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11