PlainRecalls
FDA Devices Moderate Class II Terminated

Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.

Reported: August 24, 2022 Initiated: July 19, 2022 #Z-1523-2022

Product Description

Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.

Reason for Recall

Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.

Details

Recalling Firm
Myolyn Inc.
Units Affected
45 devices
Distribution
Distribution was made to AZ, CA, CT, FL, GA, ID, IL, KY, MA, MI, MN, MO, MN, NC, NJ, NM, NY, OH, OR, TX, UT, VA, WA, and WI. There was no foreign/military/government distribution.
Location
Gainesville, FL

Frequently Asked Questions

What product was recalled?
Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.. Recalled by Myolyn Inc.. Units affected: 45 devices.
Why was this product recalled?
Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.
Which agency issued this recall?
This recall was issued by the FDA Devices on August 24, 2022. Severity: Moderate. Recall number: Z-1523-2022.