Severity
Moderate
FDA Devices recall · Reported April 1, 2020
During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK…
Karl Storz Endoscopy recalled Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope - a moderate-severity action.
Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope was recalled by Karl Storz Endoscopy in April 1, 2020. Reason: During a complaint investigation it was determined that an error was observed within the instruction manual …. Check the official notice for the remedy. Verify recall #Z-1530-2020 with the FDA Devices before acting.
The recall
Karl Storz Endoscopy issued this moderate-severity FDA Devices recall-During a complaint investigation it was determined that an error was observed within the instruction manual ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1530-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1530-2020) was formally reported on April 1, 2020, with the manufacturer initiating the action on August 27, 2019. It is classified under Moderate severity (Class II), with a current status of Ongoing. Karl Storz Endoscopy is listed as the recalling firm, operating out of El Segundo, CA. Federal records list the affected scope as 73 scopes.
The documented reason for this recall is: During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for th… Distribution data in the federal record shows the product reached: US: AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV, OUS: Canda. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
73 scopes
Related Recalls
6
3 from same agency
Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope
During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1530-2020 |
| Date reported | April 1, 2020 |
| Date initiated | August 27, 2019 |
| Recalling firm | Karl Storz Endoscopy |
| Firm location | El Segundo, CA |
| Affected scope | 73 scopes |
| Distribution | US: AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV, OUS: Canda |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 1, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.