Severity
Critical
FDA Devices recall · Reported May 19, 2021
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunctio…
St. Jude Medical, Cardian Rhythm Management Division recalled ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm As… - a critical-severity action.
ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm As… was recalled by St. Jude Medical, Cardian Rhythm Management Division in May 19, 2021. Reason: Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacem…. Check the official notice for the remedy. Verify recall #Z-1531-2021 with the FDA Devices before acting.
The recall
St. Jude Medical, Cardian Rhythm Management Division issued this critical-severity FDA Devices recall-Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacem….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1531-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1531-2021) was formally reported on May 19, 2021, with the manufacturer initiating the action on March 15, 2021. It is classified under Critical severity (Class I), with a current status of Ongoing. St. Jude Medical, Cardian Rhythm Management Division is listed as the recalling firm, operating out of Sylmar, CA. Federal records list the affected scope as 95,042 pacemakers [updated 03/20/2023].
The documented reason for this recall is: Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the… Distribution data in the federal record shows the product reached: Worldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
95,042 pacemakers [updated 03/20/2023]
Related Recalls
6
3 from same agency
ENDURITY Pulse Generator REF PM**** SN ********* St. Jude Medical Cardiac Rhythm Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Ongoing |
| Recall number | Z-1531-2021 |
| Date reported | May 19, 2021 |
| Date initiated | March 15, 2021 |
| Recalling firm | St. Jude Medical, Cardian Rhythm Management Division |
| Firm location | Sylmar, CA |
| Affected scope | 95,042 pacemakers [updated 03/20/2023] |
| Distribution | Worldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD,… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported May 19, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.