Severity
Moderate
FDA Devices recall · Reported September 11, 2024
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the Dx…
Beckman Coulter, Inc. recalled DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the qu… - a moderate-severity action.
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the qu… was recalled by Beckman Coulter, Inc. in September 11, 2024. Reason: Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash B…. Check the official notice for the remedy. Verify recall #Z-1546-2024 with the FDA Devices before acting.
The recall
Beckman Coulter, Inc. issued this moderate-severity FDA Devices recall-Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash B….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1546-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1546-2024) was formally reported on September 11, 2024, with the manufacturer initiating the action on March 5, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Beckman Coulter, Inc. is listed as the recalling firm, operating out of Chaska, MN. Federal records list the affected scope as 17 units.
The documented reason for this recall is: Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leadi… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
17 units
Related Recalls
6
3 from same agency
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1546-2024 |
| Date reported | September 11, 2024 |
| Date initiated | March 5, 2024 |
| Recalling firm | Beckman Coulter, Inc. |
| Firm location | Chaska, MN |
| Affected scope | 17 units |
| Distribution | Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported September 11, 2024. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.