FDA Devices Moderate Class II Terminated

2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K

Reported: May 6, 2015 Initiated: February 20, 2015 #Z-1551-2015

Product Description

2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K System w/o HP; 190303 2008K System w/o OLC and Diasafe; 190304 2008K System w/o OLC and Diasafe and HP; 190452 2008K System w/o OLC and w/ Diasafe Plus; 190303 2008K W/O OLC and DIASAFE W/ HEPARIN PUMP; 190304 2008K W/O OLC and DIASAFE W/O HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.

Reason for Recall

The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.

Details

Units Affected
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Distribution
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Location
Waltham, MA

Frequently Asked Questions

What product was recalled?
2008 Series Hemodialysis Machines: 2008K with the following product code and description: 190305 2008K (w/ OLC); 190371 2008K System Std (w/OLC and Diasafe Plus); 190300 2008K System Std (w/OLC and Diasafe); 190336 2008K System w/ Fifth Module Holder; 190517 2008K System w/ OLC and Diasafe Plus - Spanish; 190302 2008K System w/ OLC and Short Cab; 190306 2008K System w/ Short Cabinet; 190373 2008K System w/ Short Cabinet; 190305 2008K System w/o Diasafe; 190301 2008K System w/o HP; 190372 2008K System w/o HP; 190303 2008K System w/o OLC and Diasafe; 190304 2008K System w/o OLC and Diasafe and HP; 190452 2008K System w/o OLC and w/ Diasafe Plus; 190303 2008K W/O OLC and DIASAFE W/ HEPARIN PUMP; 190304 2008K W/O OLC and DIASAFE W/O HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213).
Why was this product recalled?
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 6, 2015. Severity: Moderate. Recall number: Z-1551-2015.