2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Product Description
2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Reason for Recall
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Details
- Recalling Firm
- Fresenius Medical Care Holdings, Inc.
- Units Affected
- 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
- Location
- Waltham, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11