Severity
Moderate
FDA Devices recall · Reported April 1, 2020
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage M…
Quidel Cardiovascular INC recalled Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0… - a moderate-severity action.
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0… was recalled by Quidel Cardiovascular INC in April 1, 2020. Reason: Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inapprop…. Check the official notice for the remedy. Verify recall #Z-1552-2020 with the FDA Devices before acting.
The recall
Quidel Cardiovascular INC issued this moderate-severity FDA Devices recall-Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inapprop….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1552-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1552-2020) was formally reported on April 1, 2020, with the manufacturer initiating the action on April 9, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Quidel Cardiovascular INC is listed as the recalling firm, operating out of San Diego, CA. Federal records list the affected scope as 123 kits.
The documented reason for this recall is: Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result… Distribution data in the federal record shows the product reached: US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
123 kits
Related Recalls
6
3 from same agency
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1552-2020 |
| Date reported | April 1, 2020 |
| Date initiated | April 9, 2019 |
| Recalling firm | Quidel Cardiovascular INC |
| Firm location | San Diego, CA |
| Affected scope | 123 kits |
| Distribution | US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported April 1, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.