Product Description

Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com

Reason for Recall

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Details

Recalling Firm
QUIDEL CARDIOVASCULAR INC
Units Affected
123 kits
Distribution
US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
Location
San Diego, CA

Frequently Asked Questions

What product was recalled?
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com. Recalled by QUIDEL CARDIOVASCULAR INC. Units affected: 123 kits.
Why was this product recalled?
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 1, 2020. Severity: Moderate. Recall number: Z-1552-2020.