Severity
Low
FDA Devices recall · Reported August 24, 2022
The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the pa…
Siemens Medical Solutions USA, Inc recalled AXIOM Sensis or Sensis Vibe Combo systems as follows: System … - a low-severity action.
AXIOM Sensis or Sensis Vibe Combo systems as follows: System … was recalled by Siemens Medical Solutions USA, Inc in August 24, 2022. Reason: The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility …. Check the official notice for the remedy. Verify recall #Z-1558-2022 with the FDA Devices before acting.
The recall
Siemens Medical Solutions USA, Inc issued this low-severity FDA Devices recall-The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1558-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1558-2022) was formally reported on August 24, 2022, with the manufacturer initiating the action on July 12, 2022. It is classified under Low severity (Class III), with a current status of Terminated. Siemens Medical Solutions USA, Inc is listed as the recalling firm, operating out of Malvern, PA. Federal records list the affected scope as 399 units U.S..
The documented reason for this recall is: The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting… Distribution data in the federal record shows the product reached: US Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cam…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified low severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Low
Affected scope
399 units U.S.
Related Recalls
6
3 from same agency
AXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology.
The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Low (Class III) |
| Status | Terminated |
| Recall number | Z-1558-2022 |
| Date reported | August 24, 2022 |
| Date initiated | July 12, 2022 |
| Recalling firm | Siemens Medical Solutions USA, Inc |
| Firm location | Malvern, PA |
| Affected scope | 399 units U.S. |
| Distribution | US Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cambodia Cameroon Can… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 24, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.