FDA Devices recall · Reported August 24, 2022

AXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well

The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the pa…

Recall #
Z-1558-2022
Affected scope
399 units U.S.
Initiated
July 12, 2022
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Siemens Medical Solutions USA, Inc recalled AXIOM Sensis or Sensis Vibe Combo systems as follows: System … - a low-severity action.

AXIOM Sensis or Sensis Vibe Combo systems as follows: System … was recalled by Siemens Medical Solutions USA, Inc in August 24, 2022. Reason: The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility …. Check the official notice for the remedy. Verify recall #Z-1558-2022 with the FDA Devices before acting.

The recall

Siemens Medical Solutions USA, Inc issued this low-severity FDA Devices recall-The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility ….

Low
severity level
399 units
affected scope
Class III
classification
August 24, 2022
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-1558-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-1558-2022) was formally reported on August 24, 2022, with the manufacturer initiating the action on July 12, 2022. It is classified under Low severity (Class III), with a current status of Terminated. Siemens Medical Solutions USA, Inc is listed as the recalling firm, operating out of Malvern, PA. Federal records list the affected scope as 399 units U.S..

The documented reason for this recall is: The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting… Distribution data in the federal record shows the product reached: US Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cam…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified low severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Low

Affected scope

399 units U.S.

Related Recalls

6

3 from same agency

Product description

AXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology.

Reason for recall

The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Low (Class III)
Status Terminated
Recall number Z-1558-2022
Date reported August 24, 2022
Date initiated July 12, 2022
Recalling firm Siemens Medical Solutions USA, Inc
Firm location Malvern, PA
Affected scope 399 units U.S.
Distribution US Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cambodia Cameroon Can…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

399 units U.S. units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-1558-2022) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
AXIOM Sensis or Sensis Vibe Combo systems as follows: System Material # AXIOM Sensis, Combo 64 6623974; AXIOM Sensis, Hemo LOW 6634633; AXIOM Sensis, Combo 32 6634641; AXIOM Sensis, EP 129 6634658; Sensis 10764561; Sensis Vibe Combo 11007642; Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 399 units U.S..
Why was this product recalled?
The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box
Which agency issued this recall?
This recall was issued by the FDA Devices on August 24, 2022. Severity: Low. Recall number: Z-1558-2022.
Where was the recalled product distributed?
Distribution: US Nationwide Distribution Foreign: Albania Algeria Angola Argentina Armenia Aruba Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Brunei Darussalam Bulgaria Cambodia Cameroon Canada Chile China Colombia Costa Rica Croatia Cuba Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guatemala Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kosovo Kuwait Latvia Lebanon Libya Luxembourg Macao Malaysia Maldives Mali Mauritius Mexico Moldova Montenegro Morocco Mozambique Myanmar Namibia Netherlands New Caledonia New Zealand Nigeria North Korea Norway Oman Pakistan Palestine, State of Panama Paraguay Peru Philippines Poland Portugal Qatar Romania Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sudan Suriname Sweden Switzerland Syria Taiwan Tajikistan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela, Bolivarian Republic of Vietnam Yemen.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-1558-2022) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported August 24, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.