FDA Devices recall · Reported May 14, 2014
Mar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: The CWP is a Central reverse osmosis system with the intended use of producing water for dialysis of multiple patients.
Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with …
Mar Cor Purification recalled Mar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: … - a moderate-severity action.
Mar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: … was recalled by Mar Cor Purification in May 14, 2014. Reason: Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. …. Check the official notice for the remedy. Verify recall #Z-1563-2014 with the FDA Devices before acting.
The recall
Mar Cor Purification issued this moderate-severity FDA Devices recall-Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. ….
- Moderate
- severity level
- 197 units
- affected scope
- Class II
- classification
- May 14, 2014
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-1563-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Mar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: The CWP is a Central reverse osmosis system with the intended use of producing water for dialysis of multiple patients.
Reason for recall
Mar Cor Purification manufactured Central Water Plant (CWP) reverse osmosis system, models 102, 104 and 106. This correction has been initiated due to a potential situation with a specific lot of the water inlet solenoid valve. The potential exists that the water inlet solenoid valve could cease to function. If this occurs, CWP will stop producing reverse osmosis water.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1563-2014 |
| Date reported | May 14, 2014 |
| Date initiated | April 24, 2014 |
| Recalling firm | Mar Cor Purification |
| Firm location | Minneapolis, MN |
| Affected scope | 197 |
| Distribution | US Nationwide Distribution - AL, AZ, AR, CA, CO, FL, GA, ID, IN, IA, KS, KY, LA, MA, MI, MN, MS, MO, MT, NH, NJ, NM, NY,NC, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WV, WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-1563-2014) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026