PowerFlow Implatable Apheresis IV Port with attachable 9.6 F ChronoFl x Open-Ended Single-Lumen Venous Catheter, Ref A710962, UDI:(01)00801741129438 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

Reported: April 1, 2020 Initiated: October 1, 2019 #Z-1579-2020

Product Description

PowerFlow Implatable Apheresis IV Port with attachable 9.6 F ChronoFl x Open-Ended Single-Lumen Venous Catheter, Ref A710962, UDI:(01)00801741129438 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

Reason for Recall

Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.

Details

Units Affected
296 Catheters
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT,VA, VT, WA, WI, WV, WY and countries of Brazil, Canada, EMEA, Korea, Mexico, New Zealand.
Location
Tempe, AZ

Frequently Asked Questions

What product was recalled?
PowerFlow Implatable Apheresis IV Port with attachable 9.6 F ChronoFl x Open-Ended Single-Lumen Venous Catheter, Ref A710962, UDI:(01)00801741129438 Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.. Recalled by Bard Peripheral Vascular Inc. Units affected: 296 Catheters.
Why was this product recalled?
Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Which agency issued this recall?