Severity
Moderate
FDA Devices recall · Reported April 23, 2025
Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of t…
Channel Medsystems, Inc. recalled Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIA… — a moderate-severity action.
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIA… was recalled by Channel Medsystems, Inc. in April 23, 2025. Reason: Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the…. Check the official notice for the remedy. Verify recall #Z-1584-2025 with the FDA Devices before acting.
The recall
Channel Medsystems, Inc. issued this moderate-severity FDA Devices recall — Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1584-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1584-2025) was formally reported on April 23, 2025, with the manufacturer initiating the action on March 20, 2025. It is classified under Moderate severity (Class II), with a current status of Ongoing. Channel Medsystems, Inc. is listed as the recalling firm, operating out of Berkeley, CA. Federal records list the affected scope as 281.
The documented reason for this recall is: Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device Distribution data in the federal record shows the product reached: U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category — 40,409 medical devices recalls on record
Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
281
Related Recalls
6
6 from same agency
Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A
Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-1584-2025 |
| Date reported | April 23, 2025 |
| Date initiated | March 20, 2025 |
| Recalling firm | Channel Medsystems, Inc. |
| Firm location | Berkeley, CA |
| Affected scope | 281 |
| Distribution | U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Keep tracking product safety across the federal recall archive.
Recall Checker
Search the full archive by product name, brand, or recall number across every agency.
Check a product →
FDA Devices recalls
Every recall issued by FDA Devices, newest first.
Browse the feed →
RecallRadar
Live feed of the latest recalls across the FDA, CPSC and NHTSA — filter by agency and severity.
View the live feed →
Rankings
The largest recalls by units affected and the most-recalled product categories.
See the rankings →
Browse by category
Find recalls by product type to spot recurring defect patterns.
All categories →
What to do next
A step-by-step guide to refunds, repairs, and returns after a recall.
Read the guide →
Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Medline Industries, LP · 2026-06-03
Abiomed, Inc. · 2026-06-03
GE Medical Systems Information Technologies Inc · 2026-06-03
Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →
Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported April 23, 2025.
Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.