Severity
Moderate
FDA Devices recall · Reported July 17, 2013
Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few s…
Toshiba American Medical Systems Inc recalled TOSHIBA Kalare Diagnostic X-Ray System K110785 - a moderate-severity action.
TOSHIBA Kalare Diagnostic X-Ray System K110785 was recalled by Toshiba American Medical Systems Inc in July 17, 2013. Reason: Firm initiated recall because when radiographic conditions with the small focus are set and the radiography r…. Check the official notice for the remedy. Verify recall #Z-1598-2013 with the FDA Devices before acting.
The recall
Toshiba American Medical Systems Inc issued this moderate-severity FDA Devices recall-Firm initiated recall because when radiographic conditions with the small focus are set and the radiography r….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1598-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1598-2013) was formally reported on July 17, 2013, with the manufacturer initiating the action on April 3, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Toshiba American Medical Systems Inc is listed as the recalling firm, operating out of Tustin, CA. Federal records list the affected scope as 153 units.
The documented reason for this recall is: Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Aut… Distribution data in the federal record shows the product reached: US Nationwide Distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
153 units
Related Recalls
6
3 from same agency
TOSHIBA Kalare Diagnostic X-Ray System K110785
Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Automatic Exposure Function (AEC) function is used. As a result, a radiographic tube current exceeding the set value flows and a brighter image than expected is acquired.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1598-2013 |
| Date reported | July 17, 2013 |
| Date initiated | April 3, 2013 |
| Recalling firm | Toshiba American Medical Systems Inc |
| Firm location | Tustin, CA |
| Affected scope | 153 units |
| Distribution | US Nationwide Distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 17, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.