Severity
Moderate
FDA Devices recall · Reported July 3, 2013
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shi…
Beckman Coulter Inc. recalled SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, whe… - a moderate-severity action.
SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, whe… was recalled by Beckman Coulter Inc. in July 3, 2013. Reason: Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition …. Check the official notice for the remedy. Verify recall #Z-1613-2013 with the FDA Devices before acting.
The recall
Beckman Coulter Inc. issued this moderate-severity FDA Devices recall-Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-1613-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-1613-2013) was formally reported on July 3, 2013, with the manufacturer initiating the action on May 28, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Beckman Coulter Inc. is listed as the recalling firm, operating out of Brea, CA. Federal records list the affected scope as 3,353 units.
The documented reason for this recall is: Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen rea… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA Nationwide and country of Canada.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
3,353 units
Related Recalls
6
3 from same agency
SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX¿ System(s), UniCel¿ DxC 600/800 System(s) and Synchron¿ Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-1613-2013 |
| Date reported | July 3, 2013 |
| Date initiated | May 28, 2013 |
| Recalling firm | Beckman Coulter Inc. |
| Firm location | Brea, CA |
| Affected scope | 3,353 units |
| Distribution | Worldwide Distribution - USA Nationwide and country of Canada. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported July 3, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.