Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.

Reported: April 8, 2020 Initiated: February 20, 2020 #Z-1616-2020

Product Description

Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.

Reason for Recall

Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

Details

Units Affected
2214 units
Distribution
Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.
Location
Andover, MA

Frequently Asked Questions

What product was recalled?
Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.. Recalled by Philips North America, LLC. Units affected: 2214 units.
Why was this product recalled?
Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00
Which agency issued this recall?
This recall was issued by the FDA Devices on April 8, 2020. Severity: Moderate. Recall number: Z-1616-2020.